Campbell-scientific CR3000 Micrologger Manual do Utilizador Página 439

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 438
Section 10. Troubleshooting
439
10.5.2 Troubleshooting Power at a Glance
Symptoms:
PossiblesymptomsincludetheCR3000programnotexecuting;
Low12VCountoftheStatustabledisplayingalargenumber.
Affected Equipment:
Batteries,charger/regulators,solarpanels,transformers
Likely Cause:
Batteriesmayneedtobereplacedorrecharged;charger/regulators
mayneedtobefixedorrecalibrated;solarpanelsortransformersmay
needtobefixedorre
placed.
Required Equipment:
Voltmeter;5kΩresistoranda50‐Ω1Wresistorforthechargingcircuit
testsandtoadjustthechargingcircuitvoltage.
10.5.3 Diagnosis and Fix Procedures
10.5.3.1 Battery Test
The procedure outlined in this flow chart tests sealed-rechargeable or alkaline
batteries in the PS100 charging regulator, or a sealed-rechargeable battery
attached to a CH100 charging regulator. If a need for repair is indicated after
following the procedure, see Warranty and Assistance
(p. 3) for information on
sending items to Campbell Scientific.
Vista de página 438
1 2 ... 434 435 436 437 438 439 440 441 442 443 444 ... 589 590

Comentários a estes Manuais

Sem comentários