Campbell-scientific CR3000 Micrologger Manual do Utilizador Página 379

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 378
Section 8. Operation
379
and arguments and the commands wherein they are used. Parameters and
arguments for specific commands are listed in the following sections.
Table 94. API Commands, Parameters, and Arguments
Parameter
Commands in which the
parameter is used
Function of parameter Argument(s)
uri
BrowseSymbols
DataQuery
ClockSet
ClockCheck
ListFiles
Specifies the data source.
source: dl
(datalogger is data
source): default,
applies to all
commands listed in
column 2.
tablename.fieldname:
applies only to
BrowseSymbols,
and DataQuery
format
BrowseSymbols
DataQuery
ClockSet
ClockCheck
FileControl
ListFiles
Specifies response format.
html, xml, json:
apply to all
commands listed in
column 2.
toa5 and tob1 apply
only to DataQuery
mode
DataQuery
Specifies range of data with
which to respond.
most-recent
since-time
since-record
data-range
backfill
p1
DataQuery
maximum number of
records (when using
most-recent argument).
beginning date and/or
time (when using
since-time ,or date-
range arguments).
beginning record
number (when using
since-record
argument).
interval in seconds
(when using backfill
argument).
integer number of
records (when using
most-recent
argument)
time in defined
format (when using
since-time ,or date-
range arguments, see
Time Syntax
(p. 380)
section)
integer record
number(when using
since-record
argument).
integer number of
seconds (when using
backfill argument).
Vista de página 378