EnviroSMART
TM
Soil Water Content Profile Probes
;Wiring: EnviroSMART / EasyAG 12 Volts (Pin 1) -------- CR10X 12 V
; EnviroSMART / EasyAG Ground (Pin 2) ------- CR10X Ground
; EnviroSMART / EasyAG Signal (Pin 3)--------- C7
; Rain Gage Pulse Out -------------------------------- CR10X P2
; Rain Gage Ground----------------------------------- CR10X Ground
;{CR10X}
;
*Table 1 Program
01: 30 Execution Interval (seconds)
;Load Scaled Frequency to Water Content Co-effecients
a = .1957
b = .404
c = .02852
;Preload missing sensor error
1: Beginning of Loop (P87)
1: 0 Delay
2: 8 Loop Count
2: Z=F x 10^n (P30)
1: -1000 F
2: 0 n, Exponent of 10
3: 1 -- Z Loc [ ES1_SF_1 ]
3: Z=F x 10^n (P30)
1: -1000 F
2: 0 n, Exponent of 10
3: 17 -- Z Loc [ ES1_VIC_1 ]
4: End (P95)
;Measure water content scaled frequencies (SF)
5: Beginning of Loop (P87)
1: 0 Delay
2: 5 Loop Count
;See Appendix E for other SDI-12 commands
6: SDI-12 Recorder (P105)
1: 0 SDI-12 Address
2: 0 Start Measurement (aM0!)
3: 7 Port
4: 1 Loc [ ES1_SF_1 ]
5: 1.0 Mult
6: 0 Offset