Campbell-scientific CS110 Electric Field Meter Manual do Utilizador Página 14

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 88
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 13
CS110 Electric Field Meter
Rugged Construction: Ultra-reliable metallic ground connection to
reciprocating shutter (no wiping contact), brushless
stepper motor, powder-coated aluminum case,
Teflon insulators, and electro-polished 316L
stainless steel used for corrosion protection of
critical exposed metallic parts
Easy Maintenance: The stator is easily removed for cleaning (proper
cleaning does not invalidate calibration).
Instrument self-checking allows maintenance to be
performed on an as needed basis. The self-checking
also monitors internal humidity, insulator
cleanliness, and power supply voltage, and verifies
that CS110 components such as the charge
amplifier and shutter open/close are functioning
properly.
Operating
Temperature Range: -25° to 50°C standard, -40° to +85°C optional
RH Range: 0 to 100% RH
Dimensions: 15.2 x 15.2 x 43.2 cm (6 x 6 x 17 in)
Mounting: Vertical pipe 1.91 to 6.35 cm OD (0.75 to 2.5 in)
Weight: 4 kg (9 lb)
6
Vista de página 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 87 88

Comentários a estes Manuais

Sem comentários