Campbell-scientific CR800 and CR850 Measurement and Control Systems Manual do Utilizador Página 39

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 38
Section 4. Quickstart Tutorial
39
Note A period-averaging sensor has a frequency output, but it is connected to a
single-ended analog input channel and measured with the PeriodAverage()
instruction (see Period Averaging
(p. 307) ).
4.1.3.3.1 Pulses Measured
Figure Pulse Sensor Output Signal Types
(p. 39) illustrates three pulse sensor output
signal types.
Figure 7: Pulse-sensor output signal types
4.1.3.3.2 Pulse-Input Channels
Table Pulse-Input Channels and Measurements
(p. 39) lists devices, channels and
options for measuring pulse signals.
Table 2. Pulse-Input Channels and Measurements
Pulse-Input
Channel
Input Type
Data Option
CRBasic
Instruction
P1, P2
High-frequency
Low-level ac
Switch-closure
Counts
Frequency
Run average
of frequency
PulseCount()
C1, C2, C3, C4,
C5, C6, C7, C8
High-frequency
Switch-closure
Low-level ac
(with LLAC4
Low-Level AC
Conversion
Module)
Counts
Frequency
Running
average of
frequency
Interval
Period
State
PulseCount()
TimerIO()
Vista de página 38