Campbell-scientific CR800 and CR850 Measurement and Control Systems Manual do Utilizador Página 275

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 274
Section 8. Operation
275
where
GainError=±(2500*0.0006)
=±1.5mV
and
OffsetError=1.5667µV+1µV=1.00mV
Therefore,
Error=GainError+OffsetError
=±1.5mV+1.00µV
=±2.50mV
Incontrast,theerrorfora500mVinputunderthesameconstraintsis
±1.30mV.ThefigureVolt
ageMeasurementAccuracy
(p.275)illustrates
thetotalerrorwithrespecttovoltagemeasurementsforthe±2500mV
range.
Figure 87: Voltage measurement accuracy (0° to 40°C)