OBS-5+ System
FIGURE 8-1. Internal components
OBS-5+ or pressure sensor malfunction.
Open unit and inspect for: 1) broken sensor wires, and 2) loose pressure
sensor connector (FIGURE 8-1).
Check sensor power by clicking
Survey and selecting all sensors; the
green LEDs should illuminate. If they do not, the sensor power circuit
may not be working.
If the depth sensor reads high and does not change, it may need to be
cleaned (see Section 9.2, Pressure Sensor).
If the sensors appear to be in working order, the digitizer or
microcontroller may be damaged. Such problems require factory service.
31