Campbell-scientific CS547 Conductivity and Temperature Probe and A547 Manual do Utilizador Página 13

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 26
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 12
CS547 Conductivity and Temperature Probe and A547 Interface
9
;This next program set will correct errors in the EC measurement resulting
;from temperature differences.
;
31: Z=X+F (P34)
1: 4 X Loc [ Temp_degC ]
2: -25 F
3: 6 Z Loc [ A ]
32: Z=X*F (P37)
1: 3 X Loc [ Ct ]
2: 100 F
3: 7 Z Loc [ lOO_Ct ]
33: Z=X*F (P37)
1: 6 X Loc [ A ]
2: nnn F Enter TC (%/
°
C) to correct cond. reading.
3: 8 Z Loc [ TC_Proces ]
34: Z=X+F (P34)
1: 8 X Loc [ TC_Proces ]
2: 100 F
3: 8 Z Loc [ TC_Proces ]
35: Z=X/Y (P38)
1: 7 X Loc [ lOO_Ct ]
2: 8 Y Loc [ TC_Proces ]
3: 9 Z Loc [ C25mScm_l ] EC corrected for temperature.
(cont.)
5.5 Output Processing
Over large ranges, EC is not linear and is best reported as samples using
instruction P70. In limited ranges, averaging (P71) measurements over time
may be acceptable. Convention requires that the temperature at the time of the
measurement be reported.
;Output processing, convention states that the temperature be reported
;with the EC measurement.
;
36: Do (P86)
1: 10 Set Output Flag High (Flag 0)
37: Sample (P70)
1: 1 Reps
2: 3 Loc [ Ct ]
38: Sample (P70)
1: 1 Reps
2: 4 Loc [ Temp_degC ]
39: Sample (P70)
1: 1 Reps
2: 9 Loc [ C25mScm_l ]
(cont.)
Vista de página 12
1 2 ... 8 9 10 11 12 13 14 15 16 17 18 ... 25 26

Comentários a estes Manuais