CS511-L Dissolved Oxygen Probe
6. Replace the membrane and its O-ring by first placing the new O-ring at the
very bottom of the membrane cavity (see FIGURE 8-4). Remove the
paper backing from a new membrane and place the new membrane on top
of the O-ring, and then place the spacer on top of the membrane. Using
the membrane tool, install the membrane lock on top of the spacer as
shown in FIGURE 8-5. Make sure the cap is upright (not sideways) when
securing the membrane lock to the spacer.
FIGURE 8-4. Proper O-ring placement
19