Appendix B. Example Programs
B.1.2 Example CR1000 Program
'CR1000 Series Datalogger
'Declare the variable for the water level measurement
Public CS470(7)
'Rename the variable names
Alias CS470(1)=Level_m
Alias CS470(2)=Level_cm
Alias CS470(3)=Level_ft
Alias CS470(4)=Pressure_mbar
Alias CS470(5)=Pressure_psi
Alias CS470(6)=Temperature_C
Alias CS470(7)=Status
'Define a data table for 60 minute maximum and minimums
DataTable(Hourly,True,-1)
DataInterval(0,60,Min,10)
Maximum(1,Level_ft,FP2,0,0)
Minimum(1,Level_ft,FP2,0,0)
Maximum(1,Temperature_C,FP2,0,0)
Minimum(1,Temperature_C,FP2,0,0)
EndTable
'Read sensor every 60 seconds
BeginProg
Scan(60,sec,1,0)
'Code for SDI-12 measurements:
SDI12Recorder(CS470,”0”,”M!”,1,0)
'Call the data table:
CallTable(Hourly)
NextScan
EndProg
B-2