AL200 ALERT2 Encoder, Modulator, and Sensor Interface
10. Troubleshooting
Problem Possible Causes
The AL200 will not transmit event-
triggered data.
• The AL200 is currently connected
to Device Configuration Utility and
data transmissions are being
suppressed.
• The event threshold is not properly
set.
• The AL200 is not currently
measuring a level that exceeds the
event threshold.
• The AL200 has not obtained an
initial GPS fix.
The Test Button is not initiating a
data transmission.
• The AL200 has not obtained an
initial GPS fix.
• The AL200 has been configured as
a datalogger peripheral and the test
button functionality has been
disabled.
The AL200 is reporting incorrect
values.
• The applied multipliers and offsets
are not correct.
• There is an issue with the attached
sensor or sensor wiring.
The radio LED is always red. • The AL200 has been configured to
keep the radio always on.
The GPS LED is on longer than the
maximum allowed on time.
• The GPS module is attempting to
obtain leap second information and
may be on for up to 13 minutes.
11. Repair
The AL200 is designed to give years of trouble-free service with reasonable
care. However, if factory repair is needed, first contact a Campbell Scientific
application engineer to obtain an RMA (Return Materials Authorization)
number. An RMA number and product-safety documents are required prior to
any repair shipments being accepted at Campbell Scientific. See the
Assistance statement at the beginning of this manual for more information.
16