INSTRUCTION MANUAL247 Conductivity andTemperature ProbesRevision: 3/96Copyright (c) 1994-1996Campbell Scientific, Inc.

Página 2 - Warranty and Assistance

247 CONDUCTIVITY AND TEMPERATURE PROBES66.4 FIELD CALIBRATIONThe cell constant is a dimensional numberexpressed in units of cm-1. The unit cm-1 issl

Página 3 - 247 Table of Contents

247 CONDUCTIVITY AND TEMPERATURE PROBES77. MAINTENANCERoutine maintenance includes thoroughlycleaning the orifice of the 247 probe with theblack nylo

Página 4

247 CONDUCTIVITY AND TEMPERATURE PROBES88.2 TEMPERATURE MEASUREMENT ERRORThe overall probe accuracy is a combination of thethermistor's intercha

Página 5 - 2. SPECIFICATIONS

247 CONDUCTIVITY AND TEMPERATURE PROBES99. DERIVING A TEMPERATURECOMPENSATION COEFFICIENT1. Place the 247 in a sample of the solution tobe measured.

Página 6 - 5. PROGRAMMING

247 CONDUCTIVITY AND TEMPERATURE PROBES1011. ELECTRICALLY NOISYENVIRONMENTSAC power lines can be the source of electricalnoise. If the datalogger is

Página 7

247 CONDUCTIVITY AND TEMPERATURE PROBES1113. 247 SCHEMATICFIGURE 11-1. 247 Conductivity and Temperature Circuit Diagram

Página 8

247 CONDUCTIVITY AND TEMPERATURE PROBES12This is a blank page.

Página 9 - 6. CALIBRATION

This is a blank page.

Página 10

Campbell Scientific CompaniesCampbell Scientific, Inc. (CSI)815 West 1800 NorthLogan, Utah 84321UNITED [email protected]

Página 11 - 8. ANALYSIS OF ERRORS

Warranty and AssistanceThe 247-L CONDUCTIVITY & TEMPERATURE PROBE is warranted byCAMPBELL SCIENTIFIC, INC. to be free from defects in materials an

Página 12

i247 Table of ContentsPDF viewers note: These page numbers refer to the printed version of this document. Usethe Adobe Acrobat® bookmarks tab for li

Página 13 - 10. INSTRUCTION 11 DETAILS

10. Instruction 11 Details...911. Electrically Noisy Environments...1012. Lon

Página 14 - 12. LONG LEAD LENGTHS

1247 CONDUCTIVITY AND TEMPERATURE PROBES1. OVERVIEW247 Probes (247-L and 247W-L) are designed formeasuring the electrical conductivity, dissolvedsoli

Página 15 - 13. 247 SCHEMATIC

247 CONDUCTIVITY AND TEMPERATURE PROBES22.2 EC SENSORElectrodes: Passivated 316 SS with DCisolation capacitors.Cell constant: Individually calibrat

Página 16 - This is a blank page

247 CONDUCTIVITY AND TEMPERATURE PROBES3FIGURE 4-1. 247 Wiring Diagram5.2 MEASUREMENT PROGRAMMING.EC Results from Instructions 5 or 6 (chosenautom

Página 17

247 CONDUCTIVITY AND TEMPERATURE PROBES411: P6 Full Bridge01: 1 Rep02: 14 250 mV fast Range(Use 500 mV fast for 21X)03: 1 IN Chan04: 1 Excite all reps

Página 18 - Campbell Scientific Companies

247 CONDUCTIVITY AND TEMPERATURE PROBES525: P34 Z=X+F01: 3 X Loc Ct02: -.00378 F03: 3 Z Loc [:Ct ]26: P94 Else27: P55 Polynomial01: 1 Rep02: 3 X